The Walter S500 spectrophotometer is equipped with a 4-position cell holder for 10mm cells, and features a large sample compartment for up to 100mm path length cells. Its many functions include basic photometric mode, quantitative analysis, multi-wavelength tests and more. The expanded key panel input and the large LCD screen display allow for more room to work with, and a clear and easy reading of absorbance, transmittance, factor, and concentration levels. The pre-aligned lamps are conveniently placed for easy replacement, while the optical parts are mounted on an aluminum foundation which provides a steady and firm base. It is ideal for students to use in college, university, and general laboratory settings.

FEATURES

  • Large LCD screen display guarantees a clear and easy reading of absorbance, transmittance, factor and concentration levels
  • User-friendly PC interface is ver simple to set-up and use via the USB port
  • Pre-aligned lamps are conveniently placed for easy replacement when needed
  • Optical parts are mounted on an aluminum foundation which provides a steady and firm base
  • Includes 4 optical glass cuvettes

  • TECHNICAL SPECIFICATIONS

    Wavelength Range 325-1000nm
    Bandwidth 4nm
    Optical System Single Beam with Littrow System (1200 lines/mm)
    Wavelength Accuracy ±0.8nm
    Wavelength Repeatability ±0.3nm
    Wavelength Resolution 0.1nm
    Photometric Range 0-200% T; -0.3-2.5 A; 0-9999 C
    Photometric Accuracy ±0.5% T (0-100% T)
    Photometric Repeatability Better than 0.3% T
    Stray Light ≤0.05% T at 340nm
    Stability ±0.002 A/hr at 500nm
    Cuvette Capability Up to 100mm pathlength
    Detector Silicon Photodiode
    Data Output USB1.0
    Light Source Halogen Lamp
    Power Requirements AC 80-250V, 50/60Hz
    Dimensions/Weight 18"W x 14"D x 8"H; 30lbs (13.5kg)