This Walter S900 spectrophotometer is equipped with a 2-position cell holder for 10mm cells, and features a large sample compartment for up to 100mm path length cells. The multi-wavelength can work with up to 32 (can be more) wavelengths. Its many functions include basic photometric mode, quantitative analysis, kinetics test, measurement and multi-wavelength tests and more. The large 9.7” LCD touch screen panel allows for more room to work with, and a clear and easy reading of absorbance, transmittance, factor, and concentration levels. Built with Android operating software and supports Wi-Fi, Bluetooth, and USB storage. The pre-aligned lamps are conveniently placed for easy replacement, while the optical parts are mounted on an aluminum foundation which provides a steady and firm base. It is ideal for students to use in college, university, and general laboratory settings.

FEATURES

  • 9.7" LCD touchscreen display guarantees a clear and easy reading of absorbance, transmittance, factor and concentration levels
  • Built with an 8-core processor, 1.7GHz, 2GB ram, 32G flash drive
  • Android operating software and supports Wi-Fi, Bluetooth, and USB storage
  • User-friendly PC interface is very simple to set-up and use via the USB port
  • Optical parts are mounted on a metal foundation which provides a steady and firm base
  • Includes 4 optical glass cuvettes and 2 quartz cuvettes

TECHNICAL SPECIFICATIONS

Wavelength Range 190-1100nm
Bandwidth 1.8nm
Optical System Double Beam with Czerny-Turner System (1200 lines/mm)
Wavelength Accuracy ±0.3nm
Wavelength Repeatability ±0.2nm
Wavelength Resolution 0.1nm
Photometric Range 0-200% T; -0.3-3 A; 0-9999 C
Photometric Accuracy ±0.3% T (0-100% T)
Photometric Repeatability Better than 0.2% T
Stray Light
≤0.05% T at 220nm & 340nm
Baseline Flatness ±0.001 A
Stability ±0.001 A/h at 500nm
Scan Speed High, medium, low; Max 1000nm/min
Display
9.7" LCD touch screen
Cuvette Capability Up to 100mm pathlength
Detector Silicon Photodiode
Data Output USB1.0 Port, WiFi, Bluetooth
Light Source Halogen & Deuterium Lamps
Power Requirements AC 80-250V, 50/60Hz
Dimensions/Weight 21.5"W x 18.5"D x 10.3"H; 46lbs (21kg)


Email a question about this item